Medical Equipment - Initial Check

What is the requirement for safety, operational and functional checks to be performed on equipment?

Any examples are for illustrative purposes only. 

Prior to initial use and following any major repair or upgrade to a fixed or portable medical device an electrical safety test is performed in accordance with NFPA 99 -2012: 10.3 Testing Requirements.  Additionally, an operational or functional test is performed to ensure that the device performs as per manufacturer specifications, in accordance with test procedures in the manufacturer’s instructions for use.

Any equipment transported between sites should be tested to ensure that the device the electrical safety and proper operation has not been compromised while in transit. 

Reference:
EC.02.04.03
NFPA 99 -2012: Chapter 10 Electrical Equipment
Last updated on March 19, 2021
Manual: Critical Access Hospital
Chapter: Environment of Care EC

If no, please comment on how we could improve this response.

If you have additional standards-related questions regarding this topic, please use the Standards Online Submission Form
Surgeons in an operating room

Additional Resources

Train your staff with a tool to quickly and efficiently assess standards compliance with our 2021 Hospital Compliance Assessment Workbook.

A group of medical professionals have a discussion in a meeting room

Hospital Webinars

Keep learning with our Hospital Breakfast Briefings Webinar Series. Get a deep dive into our standards, chapter-by-chapter, individually or as a team.

Female doctor using digital tablet in her office.

Looking for Standards?

Get key information your hospital needs to achieve and maintain continuous standards compliance and experience operational improvement.