Planning and Implementation Activities
 Effective | September 29, 2004

Blood Warmers - Alarm Check Documentation

Q: Is Pass/Fail documentation acceptable for recording blood warmer alarm checks?

A: No. While Pass/Fail documentation is commonly used for other biomedical equipment, it is not acceptable for blood warmer alarm checks. As required by standard EC.6.20 in the Comprehensive Accreditation Manual for Laboratory and Point-of-Care Testing, the actual temperature of the plates or effluent at which the alarm sounds should be recorded.